During a procedure, air was flushed through the cool point pump tubing set and into a patient's left atrium via the tacticath se ablation catheter.The bag of fluid ran out after ablation and air was detected at the pump's bubble detector.After spiking a new bag, there was still a column of air in the tubing that needed to be cleared.When going to flush the air through the tubing, it was checked that there was no contact with the patient.After this was confirmed, the cool point pump was flushed.However, there were two cool point pumps connected to the catheters; one to the tacticath se ablation catheter, the other to the advisor hd grid catheter.The incorrect stopcock on the field was turned off to the patient to flush air from the tubing, and this was not noticed until air had made its way through the cool point pump and into the tacticath se ablation catheter.Air was confirmed to be in the patient as there was an immediate st segment elevation and a pressure drop noted.An urgent cath was done with no air visible in the coronaries, therefore no intervention was done.The patient stabilized after the cath, was extubated, taken to icu, and is doing fine.
|
Corrected information: g3, h2, h6 manufacturer narrative: the results of the investigation are inconclusive since the device was not returned for analysis.Review of the device history record was not possible as the lot number is unknown.Based on the information received, the cause of the reported incident remains unknown.
|