The previous reportability decision for this event was reversed based on a retrospective review.The affected device has been requested for investigation by the manufacturer.Device was returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).Evaluation findings: the metal part at the top of the loop has parted.The ends of the parted wires are burnt.The plastic insulation is melted where the wire is exposed.There is black residue on the plastic.Signs of customer use: yes.Evidence of use: broken loop, melted insulation, black residue.Possible manufacturer's defect: yes.The defect location is the general area of the instrument and it is defined as broken by mechanical damage generally a user error.
|
It was reported that the unit broke during case.While separating the uterus from the cervix, the loop broke.There was no patient impact, and the doctor was able to complete the case with another loop from the same box.Lot is 37ja97975, which happened to be part of the last rma 200041212 that was replacing other loops.Customer will be checking to see if they have any more of these lot code loops on the shelf.
|