A field service engineer (fse) was dispatched to address the reported event.Fse confirmed the complaint by reviewing the printouts of multiple sample clot errors and noticed 4017 specimen syringe clot error had also occurred.Fse could visually see the serum trail from the sample nozzle between the test cup and the sample cup on the incubator cover indicating a clot had been aspirated.The customer had performed several shutdowns and startups prior to fse arrival.As a preventive measure, fse performed testing using eight total triiodothyronine (tt3) test cups with water as samples and no errors reoccurred.The customer validated the analyzer by successfully performing daily check, ran quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for the serial number (b)(6) from (b)(6) 2022 through aware date (b)(6) 2023.There were no similar complaints identified during the search period.The aia-360 operator's manual under section7-2: list of flags states the following: sample clot (sc flag) dispensing was cancelled because clogging was detected during sample suction.Check that the sample is free of fibrin or other substances.If the sample is not clouded, the sample nozzle may be faulty or the piping may be blocked.If this problem occurs frequently, contact the service department.The aia-360 operator's manual under section7-1: error messages states the following: (4017) spec.Sy clog detected description: specimen clog was detected.Troubleshooting: the item is not assayed.Repeat assay.The most probable cause of the reported event could not be establish, error cleared after several analyzer restart.
|