Model Number AIA-2000 ST
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2023
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was dispatched and was able to confirm and reproduce the reported issue by restarting the computer.During trouble shooting, the fse identified the computer was unable to detect the hard drive and suspected a faulty computer system.Fse replaced the computer system along with the display, cd drive, and mounting assembly.Fse ran a daily check and quality control (qc) with results within acceptable range.No further action is required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review for similar complaints were performed for the serial number (b)(6) through the aware date.There were no similar complaints identified during the search period.The most probable cause of the reported event was due to a faulty computer system.
 
Event Description
A customer reported power failure of the software application after a power outage with software error "hard drive not recognized" on the aia-2000.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000 ST
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key17386475
MDR Text Key319757530
Report Number3004529019-2023-00376
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse