Catalog Number SGC0701 |
Device Problems
Leak/Splash (1354); Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/03/2023 |
Event Type
malfunction
|
Event Description
|
This is filed to report a leak.It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) with a grade of 4.During advancement of the steerable guide catheter (sgc), the sgc would not pass through the septum.The sgc was removed and was prepared again.During preparation, air was noted in the steerable guide catheter regardless of correct preparation per instruction for use (ifu).The sgc was replaced to complete the procedure.One clip was implanted with no reported issue, reducing mr to grade 1.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated and based on the information provided, a cause for the reported failure to advance the sgc through the septum and leak/splash could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.Na.
|
|