Model Number 0998-00-3023-53 |
Device Problems
Pumping Stopped (1503); Moisture or Humidity Problem (2986)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/18/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient, the cs100 intra-aortic balloon pump (iabp) unit has excessive condensation in the balloon catheter and stopped pumping.There was no patient harm reported.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the unit.The (fse) reported that the unit shows multiple "check iab catheter" alarms.The fse confirmed the condensation removal module fan was working correctly and found a small amount of liquid in the crm.The fse then replaced the crm and the 5000 hour maintenance kit (part#d040-00-0147).The unit then passed all functional and safety tests and was released to the customer.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) unit has excessive condensation in the balloon catheter and stopped pumping.There was no patient harm reported.
|
|