Model Number 2088TC/52
Device Problems Break (1069); Failure to Capture (1081); High impedance (1291); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2023
Event Type  malfunction  
Event Description
It was reported that during an implant procedure that the atrial lead had high pacing impedance and high capture threshold during positioning.When repositioning, the stylet was difficult to advance in the atrial lead.The lead was placed in the atrium but was not capturing.The physician suspected atrial lead damage.The atrial lead was not used and the physician elected to complete the procedure with a different atrial lead.The patient condition was stable after the procedure.
 
Manufacturer Narrative
Type of investigation, investigation findings, and investigation conclusions codes updated for no product returned.
 
Manufacturer Narrative
A device history record (dhr) review was performed, and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17401440
MDR Text Key320058142
Report Number2017865-2023-36787
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502887
UDI-Public05414734502887
Combination Product (y/n)Y
Reporter Country CodeBR
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/52
Device Lot NumberP000155071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-