• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ290
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2023
Event Type  malfunction  
Event Description
The customer reported the whole image of the evis lucera elite gastrointestinal videoscope was blurry and the object could not be recognized.The image could be restored to normal outside the body.The issue occurred when preparing the scope for use in a diagnostic gastroscope procedure.There was no delay and the procedure was completed using a similar device.There were no reports of patient or user harm associated with this event.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation found the following: due to wear of angle wire, bending angle in upwards direction does not meet the standard value; switch1 has discoloration; the angle wires become stretched; the entire image is blurry (the subject cannot be recognized and image does not return); no up angle due to broken air-wire; broken protector; the adhesive around objective lens was cracked; and fog in the image after cold and hot water testing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The g2 field was corrected based on the information available at the time of the initial submission.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 4 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The event can be detected by handling the device in accordance with the following section of the instructions for use: "inspection of the endoscopic image." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17403989
MDR Text Key319846954
Report Number9610595-2023-10672
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-HQ290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-