• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804400-23
Device Problems Difficult to Remove (1528); Migration (4003)
Patient Problems Bradycardia (1751); Low Blood Pressure/ Hypotension (1914); Device Embedded In Tissue or Plaque (3165)
Event Date 07/11/2023
Event Type  Injury  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that the procedure was to treat a lesion located in the ostial right coronary artery (rca).Post- deployment of the 4.0x23mm xience skypoint stent, there was slight resistance removing the stent delivery system.The stent was pulled half way into the aorta and only a portion was still in the proximal rca.This was only observed on post imaging and was not realized during removal of the delivery system.The patient's heart rate was elevated and a balloon pump was put in and medications were administered [dopamine] as treatment.The patient was emergently transferred to the cath lab across the street.[st.Joseph atlanta].Additional follow-up indicated treatment included embedding the stent with a flare balloon and implanting an additional stent with great results.The patient did well.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to the operational context of the procedure.The reported patient effect of hypotension is listed in the xience skypoint everolimus eluting coronary stent systems instructions for use as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined; however, the subsequent treatments appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17419965
MDR Text Key320020763
Report Number2024168-2023-08214
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648233494
UDI-Public08717648233494
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/20/2023
Device Catalogue Number1804400-23
Device Lot Number1121641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-