During examination, the customer's reported issue was confirmed; a pin hole was observed in the instrument channel.The channel was blocked due to the foreign material found at this area.Examination also showed the bending section cover's adhesive was chipped, cracked and partially detached; the control unit was corroded; the light guide cover glass was scratched; the ultrasound transducer was corroded; the objective lens and light guide lenses were both scratched; and the connecting tube was scratched.Functional testing showed the up direction's bending angle was out of specifications and the up/down knob had excess play; both issues occurred due to a worn angle wire.In addition, the ultrasonic image was incomplete due to probe damage.The investigation is ongoing.A supplemental report will be submitted upon completion of investigation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the foreign object could not be removed due to physical damage of the device.A pinhole was confirmed in the forceps channel.It was unclear whether the reprocessing was performed according to the instructions for use (ifu).The event may be prevented by following the descriptions in the following items in the instructions for use (ifu): [instruction manual evis lucera ultrasonic gastrovideoscope olympus gf type uct260] chapter 6 application and conditions of cleaning, disinfection and sterilization.Chapter 7 cleaning, disinfection and sterilization procedures.1) "when the curved portion of the endoscope is greatly curved, the force required to insert/withdraw the instrument increases.If it is difficult to insert/remove the treatment instrument, return the bending angle of the bending part to a point where insertion/removal can be done without difficulty.Forcible insertion/removal may damage the forceps channel and the instrument." 2) "make sure that the tip of the treatment instrument is closed or retracted into the sheath, and then slowly insert or withdraw it into the forceps plug.Also, do not open the tip of the treatment instrument or pull it out of the sheath while it is being inserted into the forceps channel of the endoscope.The forceps channel and treatment tools may be damaged." olympus will continue to monitor field performance for this device.
|