• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Device Alarm System (1012); Failure to Deliver (2338); No Flow (2991)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that as soon the device was turned on the over temperature alarm turns on and the unit stops circulating the liquid.Patient involvement unknown.
 
Manufacturer Narrative
Evaluation codes: updated device evaluation: one device received for investigation.Visual inspection noted: device received very dirty, power switch is damaged and connected to the old style printed circuit board (pcb), water tank is stained with rust, corroded quick connect, pole clamp is discolored, microswitch was stuck onto plate clip, front cover and enclosure have multiple deep scratches.Functional testing found the device started alarming immediately.Root cause was attributed to a faulty pump.Failed water pump no longer recirculating water due to wear of age.Device manufactured in 2005 and first time in for service.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped or returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17430997
MDR Text Key320157333
Report Number3012307300-2023-07652
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer07/19/2023
Was the Report Sent to FDA? No
Date Manufacturer Received09/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-