SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C45101314J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 07/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other text: d4: lot number, expiration date, udi number and h4: device manufacture date is unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that there was leak from the anesthesia circuit.Patient involvement unknown.
|
|
Manufacturer Narrative
|
Two photos of the affected device and one sample device were received for investigation.During visual inspection, embedded resin was detected on the device tubing.The sample was functionally tested, and leakage was observed from the tubing.No product lot number was provided, therefore, no review of manufacturing device history records could be conducted.Based on the results of the inspection and testing, the investigation confirmed the reported issue.The root cause of the observed condition was attributed to an error during the manufacturing process.Complaint information will continue to be monitored and actions assigned as required.
|
|