H10: through follow-up communication with livanova field service representative, it was learned that issue was with the flow sensor, not with the scp.In detail, flow sensor was not reading flow.Therefore, it was put out of use by the customer and replaced with another one by livanova technician.Subsequent functional verification testing was completed without issues and the unit was put to service.Flow sensor not reading the flow is always detectable before patient use, therefore the event has been re-assessed as not reportable.
|