Model Number 16-02-85 |
Device Problems
Erratic or Intermittent Display (1182); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/10/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t device had a display malfunction during procedure: it came on and then went out, without came back on.There was no patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in colorado springs, colorado.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The unit was tested, upon power up the on/off switch, power was shown.However there was no display powered on or fan running.In order to fix it, service kit mains board was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Field service eng repair the device replacing dedicate electronic board ul507.All functional verification test completed with positive result.The device has returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: a complaints database review was performed and no other similar issue has been reported since unit installation in 2016.Based on livanova technical intervention, the root cause of the reported issue can be traced back to a faulty mains board, most likely due to the wearing of its electronics.
|
|