DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cardiosave intra-aortic balloon pump (iabp) unit had a crack display cover.There was no patient involvement reported.
 
Manufacturer Narrative
Additional info: contact person(name: (b)(6).It was reported that the cardiosave intra-aortic balloon pump (iabp) had console damage / malfunction.There was no patient involved or adverse event reported.A getinge fse was dispatched to evaluate the unit.The fse replaced top cover to resolve the issue.Device passed safety and functional tests and was returned to customer and cleared for customer use.The failure analysis and testing dept.Received part with a reported unit failure of a crack on the top cover.The fat performed a visual inspection and found the part to have a thin crack.Fat was able to verify the reported issue.Retaining the part in the failure analysis and testing department per procedure number 0002-07-d008 rev.Ap.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17436375
MDR Text Key320171837
Report Number2249723-2023-03460
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/10/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-