LIVANOVA DEUTSCHLAND S5 ROLLER PUMP 150 (S5 RP150); CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-80-00 |
Device Problem
Decreased Pump Speed (1500)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/04/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6), germany.During maintenance and repair activity done by a livanova field service representative, in addition to encoder replacement, s5 system was flashed to a new firmware (r324).Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that rotary encoder of a s5 roller pump was replaced during maintenance activity since it gave an error message (error setpoint encoder).It cannot be excluded that pump speed was not adjustable.There was no patient involvement.
|
|
Manufacturer Narrative
|
As per follow-up communication, error message "shaft angle encoder fault pump 8" was displayed.It signals that the pump speed can no longer be adjusted correctly.As troubleshooting the shaft encoder (#96-414-506) was replaced.Functional test passed positively and unit was returned to service.Complaint database review revealed that unit was installed in 2018 and no further similar issues are reported.Based on the outcome of service activity, it cannot be ruled out that the most likely root cause of the reported event is a defective shaft encoder.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See initial report.
|
|