LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-85 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/11/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in flushing, new york.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue, when the it was turned on.In order to solve the issue, stirrer motor and distribution board were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message related to stirring mechanism blocked / too slow (patient side) and the temperature dropped.The issue occurred during procedure.There was no patient injury.
|
|
Manufacturer Narrative
|
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2015.Based on the available information and taking into account the review of similar events, it cannot be excluded that because of rusted/corroded/damaged motor bearings, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, the pump was no longer rotating freely and required a power overload to work, which could have led to an overcurrent that ultimately caused damages to the distribution board.
|
|
Event Description
|
See initial report.
|
|