• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2023
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in flushing, new york.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue, when the it was turned on.In order to solve the issue, stirrer motor and distribution board were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message related to stirring mechanism blocked / too slow (patient side) and the temperature dropped.The issue occurred during procedure.There was no patient injury.
 
Manufacturer Narrative
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2015.Based on the available information and taking into account the review of similar events, it cannot be excluded that because of rusted/corroded/damaged motor bearings, likely due to environmental conditions, as water infiltration, humidity, condensation or high temperatures, the pump was no longer rotating freely and required a power overload to work, which could have led to an overcurrent that ultimately caused damages to the distribution board.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17442752
MDR Text Key320234667
Report Number9611109-2023-00374
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-