Model Number Ø7X 30MM BC IF SCRW, VENTED
Device Problems Break (1069); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/15/2023
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
It was reported that during an anterior cruciate ligament surgery the screw fractured while being inserted.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update avoe (b)(6) 2023.It was confirmed that no broken parts remained inside the patient.
 
Manufacturer Narrative
The complaint was confirmed.One unpackaged ar-4030c-07 serial/batch (b)(6) was returned for investigation.The returned interference bio screw was visually inspected and noted that it was broken.Damage was observed inside the screw.The inside geometry was lost.The method of bone preparation employed during the procedure, as well as the bone quality encountered, was not provided.The most likely cause is continually applying torque when the implant is seated and /or not advancing.Per dfu-0111 at revision 1.G.Precautions.6.Bio-absorbable interference screw only: it is important to completely seat the screwdriver to prevent potential stripping of the hex and/or screw fracture during insertion or removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ø7X 30MM BC IF SCRW, VENTED
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17445925
MDR Text Key320385638
Report Number1220246-2023-07393
Device Sequence Number1
Product Code MAI
UDI-Device Identifier00888867277649
UDI-Public00888867277649
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K180662
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberØ7X 30MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4030C-07
Device Lot Number15003347
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-