The customer reported to olympus, the evis lucera ultrasound gastrovideoscope had an unclear image and the balloon when inflated, deflates on its own.The issue was found during the procedure and was completed with the same device.There were no reports of patient harm.
|
The device was returned to olympus for evaluation, and the customer¿s allegation was not confirmed.The evaluation found that the no image and balloon inflation issue were unable to be confirmed.There were additional findings as follows: there was an insufficient angle on the angle wire pointing up, the up/down knob is out of standard value, the distal sheath rubber¿s adhesive is detached, the air/water cylinder has paint peeling, there is a rubber scratch on the switch 1, the objective lens has adhesive peeling, and the connecting tube has a scratch.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|