• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT260
Device Problems Poor Quality Image (1408); Unintended Deflation (4061)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the evis lucera ultrasound gastrovideoscope had an unclear image and the balloon when inflated, deflates on its own.The issue was found during the procedure and was completed with the same device.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was not confirmed.The evaluation found that the no image and balloon inflation issue were unable to be confirmed.There were additional findings as follows: there was an insufficient angle on the angle wire pointing up, the up/down knob is out of standard value, the distal sheath rubber¿s adhesive is detached, the air/water cylinder has paint peeling, there is a rubber scratch on the switch 1, the objective lens has adhesive peeling, and the connecting tube has a scratch.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17446477
MDR Text Key321430030
Report Number3002808148-2023-07889
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2023
Was the Report Sent to FDA? No
Date Manufacturer Received07/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-