(b)(4).The distributor reported on behalf of a healthcare facility in china, that the gas outlet port of a rd900 neopuff infant resuscitator was broken.Method: the complaint rd900 neopuff infant resuscitator was returned to the fisher & paykel healthcare (f&p) service center in china, where it was inspected by a trained f&p service technician.Results: inspection of the rd900 neopuff infant resuscitator revealed that the gas outlet port was broken.Conclusion: we are unable to determine the cause of the reported event.However, the most likely cause of the reported event is due to phsical impact.Device background: the neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff would have met the requirements at the time of production.The neopuff technical manual states the following: - dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.
|