• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL Back to Search Results
Model Number RD900
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The distributor reported on behalf of a healthcare facility in china, that the gas outlet port of a rd900 neopuff infant resuscitator was broken.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The distributor reported on behalf of a healthcare facility in china, that the gas outlet port of a rd900 neopuff infant resuscitator was broken.Method: the complaint rd900 neopuff infant resuscitator was returned to the fisher & paykel healthcare (f&p) service center in china, where it was inspected by a trained f&p service technician.Results: inspection of the rd900 neopuff infant resuscitator revealed that the gas outlet port was broken.Conclusion: we are unable to determine the cause of the reported event.However, the most likely cause of the reported event is due to phsical impact.Device background: the neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff would have met the requirements at the time of production.The neopuff technical manual states the following: - dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Type of Device
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17447085
MDR Text Key321436597
Report Number9611451-2023-00722
Device Sequence Number1
Product Code BTL
UDI-Device Identifier09420012412201
UDI-Public(01)09420012412201(10)2101393501(11)201118
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number2101393501
Was Device Available for Evaluation? No
Date Manufacturer Received07/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-