Model Number 71005
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/22/2022
Event Type  Injury  
Manufacturer Narrative
Per the physician, there is no evidence of component migration.Physican states the ipg was implanted at an inappropriate depth and the patient's issues were resolved with placing the new ipg at a more optimal location.Patient currently receives pain relief therapy.
 
Event Description
Patient was implanted with the nalu spinal cord stimulator system on (b)(6)2022.After activation, the patient reported lack of pain relief and difficulty using the system.Per the physician, imaging shows that the implantable pulse generator (ipg) was placed at a depth that was not optimal for maintaining connectivity with the external therapy discs.Surgical revision was performed on (b)(6)2022.During the procedure the existing ipg was replaced and the new component was placed in a more optimal position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17447317
MDR Text Key320281638
Report Number3015425075-2023-00154
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00812537036137
UDI-Public01008125370361371121113017241130
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71005
Device Catalogue Number71005
Device Lot NumberA03155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-