MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD SENSATION 7FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD SENSATION 7FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0434
Device Problems Difficult to Remove (1528); Unraveled Material (1664)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 07/11/2023
Event Type  Injury  
Event Description
It was reported that resistance was met while inserting the guidewire.The scrub practitioner advanced the guidewire provided in the kit with the catheter instead of holding steady.The intra-aortic balloon (iab) coiled inside the patient while inserting over the guidewire.The guidewire then could not be removed easily so the iab and guidewire both had to be removed.The femoral artery suffered some damage during the whole process and had to be repaired by a vascular surgeon.The patient was then placed on extracorporeal membrane oxygenation (ecmo) and had since been weened off and recovering well in the hospital.
 
Manufacturer Narrative
Department: perfusion.Event site postal code: m23 9lt.The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.Complaint record id # (b)(4) h3 other text : device not returned.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint#: (b)(4).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION 7FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key17449698
MDR Text Key320286189
Report Number2248146-2023-00486
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567106779
UDI-Public10607567106779
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0684-00-0434
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexMale
-
-