MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Manufacturer Narrative
H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.Ring transformer and ac mains board were replaced without solving the issue.Besides the cardioplegia bridge, all original parts were re-installed in the machine, without improvements.It cannot be ruled out that the issue was due to the compressor or the cooling system.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t did not cool patient side, when cardioplegia one was running.The issue occurred during procedure.There was no patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
According to complaints database review no other similar issue have been submitted for this unit since device installation in 2014.The root cause of the reported issue has been traced back to a faulty compressor.Taking into account the manufacturing year of the device,the wearing of electro-mechanical device, that can be originated by the specific use condition at customer site, may have contributed to the event, however there is no concerning trend for this kind of failure.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.This report was due on november 24, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 19, 2023, before this report was ready to submit.The report was prepared and submitted following restoration of the livanova systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17451006
MDR Text Key320319710
Report Number9611109-2023-00381
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-