H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.Ring transformer and ac mains board were replaced without solving the issue.Besides the cardioplegia bridge, all original parts were re-installed in the machine, without improvements.It cannot be ruled out that the issue was due to the compressor or the cooling system.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
According to complaints database review no other similar issue have been submitted for this unit since device installation in 2014.The root cause of the reported issue has been traced back to a faulty compressor.Taking into account the manufacturing year of the device,the wearing of electro-mechanical device, that can be originated by the specific use condition at customer site, may have contributed to the event, however there is no concerning trend for this kind of failure.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.This report was due on november 24, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 19, 2023, before this report was ready to submit.The report was prepared and submitted following restoration of the livanova systems.
|