Catalog Number LR-EVN-13.0-RL |
Device Problems
Entrapment of Device (1212); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 08/02/2023 |
Event Type
Injury
|
Event Description
|
Reported by the dm on behalf of the customer via email - the 13fr evolution was being utilized to extract the right atrial lead.The right atrial lead was freed from the heart, however, the right ventricular lead experienced lead wrapping in the process.The 13fr evolution was then found to be stuck in the heart and caught on the right ventricular lead.The choice was made to convert to an open procedure, where it was found that the right ventricular lead had gotten stuck in the 13fr evolution.Both leads were removed from the device via the open procedure.And the device was removed from the body.
|
|
Manufacturer Narrative
|
G5 ¿ pma/510(k): k141148.The event is currently under investigation.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available or upon completion of investigation.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.G5 ¿ pma/510(k): (b)(4).The device was not returned for the complaint, therefore a physical investigation could not be performed and the customer's complaint could not be confirmed, other than by the customer's testimony.The complaint/event that was entered and reported within trackwise: "device stuck in heart and caught on right ventricular lead." the device history record (dhr) was unable to be viewed due to the lot number specific to this complaint was unknown.This complaint mode is being monitored, tracked and trended per the cvi post market surveillance and complaint handling processes.A risk assessment will be performed and documented in the complaint summary tab in trackwise.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Reported by the district manager on behalf of the customer via email - the 13fr evolution was being utilized to extract the right atrial lead.The right atrial lead was freed from the heart, however, the right ventricular lead experienced lead wrapping in the process.The 13fr evolution was then found to be stuck in the heart and caught on the right ventricular lead.The choice was made to convert to an open procedure, where it was found that the right ventricular lead had gotten stuck in the 13fr evolution.Both leads were removed from the device via the open procedure.And the device was removed from the body.
|
|