MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367955
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  malfunction  
Event Description
It was reported that while using bd vacutainer® sst¿ ii advance plus blood collection tubes that there was a broken lid/ cap.The following information was provided by the initial reporter: the outpatient blood collection team of (b)(6) hospital, found 367955 yellow cranial defects when drawing blood.
 
Manufacturer Narrative
H.6 investigation summary: bd had not received samples, but one (1) photo was provided for investigation.The photo was reviewed and the indicated failure mode for molding defect/short shots was observed.Additionally, one hundred (100) retention samples from bd inventory were evaluated by visual examination and the issue of molding defects/short shots was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17455571
MDR Text Key321087949
Report Number9617032-2023-01037
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679558
UDI-Public50382903679558
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2023
Device Catalogue Number367955
Device Lot Number2175209
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-