Catalog Number 367841
Device Problems Component Incompatible (1108); Short Fill (1575)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/31/2023
Event Type  malfunction  
Manufacturer Narrative
H.6 investigation summary: bd had not received samples, but two (2) photos were provided for investigation.The photos were reviewed and the indicated failure modes for cocked stopper and underfill were observed.Ninety (90) retention samples from bd inventory were evaluated by visual examination and the issue of cocked stopper was not observed.Additionally, ten (10) retention samples from bd inventory were evaluated by functional testing and the issue of underfill was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure modes of cocked stopper and underfill based on the returned photos.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Event Description
It was reported that bd vacutainer® k2 edta (k2e) 3.6mg blood collection tubes underfill and slight slanted cap was observed.2 tubes were affected.The following information was provided by the initial reporter: users observed an underfill in the 367841 tube and attributed to slight slanting of tube cap.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 3.6MG BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17455780
MDR Text Key320581079
Report Number1917413-2023-00726
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903678414
UDI-Public50382903678414
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/30/2023
Device Catalogue Number367841
Device Lot Number2166018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-