All available information was investigated, and the reported m-cable break, inability to curve with the m/l knob, and inability to straighten with the m/l knob were confirmed via returned device analysis.The reported noise could not be replicated in a testing environment.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported noise and inability to curve and straighten with the m/l knob are cascading events of the reported m-cable break.The broken m-cable appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
|