• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C9960KP
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It as reported that the two (2) large volume multirate infusors did not infuse the medicine.The device did not purge, nor did it perform forced purging with three-way valves and a syringe.The issue was identified during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Correction b5: change quantity to one (1), previously reported as two (2) devices.H4: the device was manufactured from june 30, 2022 ¿ july 1, 2022.H10: the device was received for evaluation.A visual inspection with the naked eye was performed which did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed and the results were found to be within the product specification range.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17458247
MDR Text Key320390271
Report Number1416980-2023-03919
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081496
UDI-Public(01)00085412081496
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K011317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C9960KP
Device Lot Number22F038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received09/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-