• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 550; CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS Back to Search Results
Model Number 550AHCT
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's subsidiary, the bpm would slowly revert to reading values between two and three, while the correct values were between four and five.This was tested using manual blood gas results.
 
Event Description
It was reported that during use of the device for cardiopulmonary bypass (cpb), the blood parameter monitor (bpm) potassium (k+) levels were reading incorrectly.The bpm was recalibrated mid case, it worked for a small period of time, before the readings were incorrect again.Manual blood gas analysis was used for the remainder of the case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) connected the blood parameter monitor (bpm) to lab use only (luo) testing equipment.The monitor was set to operate mode with a standard reference sensor (srs) and cuvette attached to the respective probes.The monitor was allowed to run in operate mode for approximately one hour.During this time several in vivo adjustments were made to the potassium (k+) value all with the correct response.The reported behavior was not present during the evaluation.
 
Manufacturer Narrative
The reported complaint was not confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Per clinical review on 12jul2023, the team experienced a problem with their blood parameter monitor (bpm) during cardiopulmonary bypass (cpb), described as potassium (k+) levels reading incorrectly, with no delay, no blood loss, and successful completion of the procedure.
 
Manufacturer Narrative
The service repair technician (srt) found the arterial blood parameter monitor (bpm) probe to exceed alert limits.The arterial bpm probe was replaced as the most likely cause of the reported issue.The unit operated to the manufacturer's specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Type of Device
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key17459862
MDR Text Key320525529
Report Number1828100-2023-00214
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00886799001790
UDI-Public(01)00886799001790(11)200707
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K182110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number550AHCT
Device Catalogue Number550AHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
BLOOD GAS ANALYZER
-
-