The reported event could not be confirmed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.The identification of the device could be confirmed based on the catalog # and lot # marked.The received target device gamma 3 shows traces of a frequent and intense use, with multiple usage marks observed all over the device.The printings on the target device were turned from white to fawn and are partly worn; an indication for a high level of sterilization cycles during 15 years of use.The target device returned is an of old design version.A simulation of a surgical function test was performed on the returned device.The drills passed the nail's holes with no contact.The alleged mis-targeting could not be reproduced.Therefore, the function of the target device was still given.Based on investigation, the root cause was attributed to a user related issue.A deficiency of the returned device was not verified.
|