• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. MEDEX TRANSTAR SINGLE MONITORING KIT; PROBE, BLOOD-FLOW, EXTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. MEDEX TRANSTAR SINGLE MONITORING KIT; PROBE, BLOOD-FLOW, EXTRAVASCULAR Back to Search Results
Catalog Number MX9505T
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Manufacturer Narrative
Other text: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Event Description
It was reported that when using the device with a patient, it was found that the sensor value drifted and could not be normally used.After several calibrations, the value would always drift upwards and was thus replaced with a new sensor.No patient adverse event.The product's usage was stopped on (b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX TRANSTAR SINGLE MONITORING KIT
Type of Device
PROBE, BLOOD-FLOW, EXTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6250 shier rings road
dublin OH 43016
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17464666
MDR Text Key320724553
Report Number3012307300-2023-07824
Device Sequence Number1
Product Code DPT
UDI-Device Identifier10351688503992
UDI-Public10351688503992
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K942377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMX9505T
Device Lot Number4299812
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-