• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; N/A Back to Search Results
Catalog Number A1114
Device Problem Mechanical Problem (1384)
Patient Problem Insufficient Information (4580)
Event Date 07/07/2023
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the torque screw on mayfield infinity skull clamp (a1114) does not exert pressure, and the force indicator of the adjustment screw does not allow to visualize the value of the force applied in increments of 20 lbs.The failure occurred in surgery with a patient, which caused an adverse event.However, no surgical delay nor patient injury reported.
 
Event Description
N/a.
 
Manufacturer Narrative
The mayfield infinity skull clamp (a1114) was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined; however, based on the reported complaint, probable root cause is damage to the torque screw or an out of specification hastalloy spring.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17466122
MDR Text Key320517888
Report Number3004608878-2023-00141
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/15/2023
Date Device Manufactured04/17/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-