H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.H10: d4 (expiration date: 11/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
H10: as this malfunction is considered one event identified by the dealer; only one mdr report will be submitted for the reported quantity affected of four for this event.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Additionally, sterilization records were reviewed and this lot met all sterilization release criteria.Investigation summary: four presto devices were returned for evaluation.The devices were all received in a presto white shipping box.Damage to the shipping box was noted, however this appeared to be a result of shipping.Yellow staining was noted to the outer white yellow shipping box.The devices were removed from the box, and no evidence of contamination was noted to the device trays or sterile barrier.No contamination was seen on the physical devices.The seals of all four samples were noted to be intact and correctly sealed.Ftir and sem/eds analysis was performed on the yellow staining.The results of this testing were compared to manufacturing records, and no match to the material was able to be identified.The source of the staining could not be determined.Four electronic photos were provided for review.Two photos show a presto device tray partially inside a white shipping box and cardboard box.A yellow staining can be seen on the white shipping box, and on the sterile tray itself.The sterile tray appears to be peeled where the staining was present on the tyvek layer.One photo shows the white shipping box outside of the carboard box.The yellow staining can be seen on the outside of the box.The last photo shows the carboard shipping box.No staining or damage can be seen to the outer carboard box.Any labelling visible matches the reported information in trackwise.Therefore, the investigation is confirmed for the reported contamination, as an unknown yellow staining was noted to the outer white shipping box.However the investigation is unconfirmed for the unsealed packaging, as no evidence of damage to the sterile barrier was noted to the four device samples.The definitive root cause for the reported contamination and unsealed packaging could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 11/2025), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|