Zoll medical united kingdom evaluated the device and the device performed to specification.The customer's report was not observed during evaluation of the device using the customer's returned accessories.The device passed functional stress testing, analyze, sync, and shock testing without duplicating the report.The device was recertified and returned to the customer.Review of the clinical log confirmed the device shocked on a t-wave.Prior to the shock, there is evidence of inconsistent r wave detection leading up to the shock.There was low amplitude, high frequency artifact on the ecg trace during the shock which could have affected the device's ability to detect an r wave.It is not known what caused the artifact, however, potential sources include motion artifact from the patient or the cable, poor connection between the pads and the patient, incorrect lead application, or the patient's inherent ecg waveform under the selected lead.Per the r series operator's guide "verify that markers are clearly visible on the monitor and their location is appropriate and consistent from beat to beat.If necessary, use the lead and size buttons to establish settings that yield the most consistent sync marker pattern." the device's ecg rhythm analysis and internal sync detect is performed by a sophisticated algorithm that can be affected by artifact that can cause the device to misidentify the patient's r wave.The operator's guide also states "inappropriate defibrillation or cardioversion of a patient may precipitate ventricular fibrillation, asystole, or other dangerous arrhythmias." no trend is associated with reports of this type.
|