MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE LIGATING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE LIGATING DEVICE Back to Search Results
Model Number HX-400U-30
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2023
Event Type  Injury  
Event Description
An olympus representative reported on behalf of the customer that the snare loop of the single use ligating device could not be released during a polypectomy.In addition, the wire was bent and broke, so the operator struggled with the strength of the slider.The wire was cut with a nipper according to the instructions for use and the ligature part was cut using a loop cutter.The procedure was completed with a similar device.The device was inspected prior to use with no issues noted.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
E1 - complete establishment name: local independent administrative institution general hospital national health insurance (b)(6) hospital.The suspect device has not been returned to olympus.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer¿s reported complaint was confirmed.The device evaluation findings are as follows: the sheath was severed at the root of the handle portion.The operating pipe of the slider was broken.The broken portion presented a fracture shape caused by a mechanical load.The broken portion of the sheath presented a shape that was possibly severed by a tool.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, the root cause could not be determined.However, the reported event (loop failed to release) was likely due one of the following mechanisms: mechanism #1: 1) the loop was surrounding the body tissue, and it was temporarily ligated by pulling the slider.2) the tube sheath was pushed out, and the distal end of the coil sheath went into the tube sheath.3) an attempt was made to detach the loop in the state of above (#2).Therefore, the loop detached from the hook in the tube.4) while the hook was extending from the coil sheath, the loop moved towards the proximal side and went into the coil sheath.5) the hook was pulled.This caused the hook and the loop to retract into the coil sheath together.As a result, the loop and the hook got stuck inside the coil and could not move.6) since the loop and the hook got stuck together inside the coil, the loop did not detach when the slider was pulled.7) the slider was forcefully operated in state of ¿6¿ description.This had caused the operating pipe to bend and to break.Mechanism #2: 1) the sheath was bent near the handle.2) the operating wire could not move because sliding resistance between the sheath and the operating wire increased.3) the operating pipe deformed and broke because the slider was forcefully operated.4) due to the above, the loop could not detach from the hook.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown in ¿equipment to be used in an emergency¿ in this manual.¿ ¿do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to remove.In this case, refer to section 12, ¿emergency treatment¿ and as shown in ¿equipment to be used in an emergency¿ in this manual.¿ ¿do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to remove.In this case, refer to section 12, ¿emergency treatment¿ and as shown in ¿equipment to be used in an emergency¿ in this manual.¿ ¿never use excessive force to operate the instrument.This could damage the instrument.¿ ¿straighten out the portion of the instrument that extends from the biopsy valve.¿ this supplemental report includes an update to d9 and h3.Also, new information has been added to d4 and h4.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE LIGATING DEVICE
Type of Device
SINGLE USE LIGATING DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17470200
MDR Text Key320511742
Report Number9614641-2023-01097
Device Sequence Number1
Product Code FHN
UDI-Device Identifier04953170368615
UDI-Public04953170368615
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHX-400U-30
Device Lot Number34K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCOPE WITH UNKNOWN MODEL NUMBER
Patient Outcome(s) Required Intervention;
-
-