• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC HEMOPRO POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number VH-3010
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2023
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, t.W.Power supply s/n (b)(6), was overheating.They were able to finish the case without changing anything.No issues with vein, blood loss or burns from the "overheating".Biomed has checked the generator without finding any issues or faulty operating.They said it could possibly be the vh-4030 cord , but the cord was put back in circulation as they had not changed it during the procedure.Power setting was at "3" and no other issues during the harvest.Delay was 8-10 minutes.No patient or vein issues.
 
Manufacturer Narrative
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, t.W.Power supply s/n: (b)(6), was overheating.Device was put back into circulation.They were able to finish the case without changing anything.No issues with vein, blood loss or burns from the "overheating".Biomed has checked the generator without finding any issues or faulty operating.Power setting was at "3" and no other issues during the harvest.Delay was 8-10 minutes.No patient issues.
 
Manufacturer Narrative
Trackwise id#: (b)(4).The reported device is an oem device.The certificate of conformance was reviewed for the serial#: (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text: device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOPRO POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17470241
MDR Text Key320782725
Report Number2242352-2023-00660
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-3010
Device Catalogue NumberVH-3010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/20/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2016
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK.; UNKNOWN.
-
-