Model Number VH-3010 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/14/2023 |
Event Type
malfunction
|
Event Description
|
The hospital reported that during an endoscopic vein harvesting procedure, t.W.Power supply s/n (b)(6), was overheating.They were able to finish the case without changing anything.No issues with vein, blood loss or burns from the "overheating".Biomed has checked the generator without finding any issues or faulty operating.They said it could possibly be the vh-4030 cord , but the cord was put back in circulation as they had not changed it during the procedure.Power setting was at "3" and no other issues during the harvest.Delay was 8-10 minutes.No patient or vein issues.
|
|
Manufacturer Narrative
|
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
|
Event Description
|
The hospital reported that during an endoscopic vein harvesting procedure, t.W.Power supply s/n: (b)(6), was overheating.Device was put back into circulation.They were able to finish the case without changing anything.No issues with vein, blood loss or burns from the "overheating".Biomed has checked the generator without finding any issues or faulty operating.Power setting was at "3" and no other issues during the harvest.Delay was 8-10 minutes.No patient issues.
|
|
Manufacturer Narrative
|
Trackwise id#: (b)(4).The reported device is an oem device.The certificate of conformance was reviewed for the serial#: (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.H3 other text: device not returned.
|
|