Catalog Number 777414 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during placement of the urethral stent by the director and professor on (b)(6) 2023, the product was unpacked after the inspection showed that the outer package was intact.The guide wire could not be inserted.The head nurse was asked to use another new box of the urethral stent to complete the procedure.
|
|
Manufacturer Narrative
|
The reported event is inconclusive as no sample was returned for evaluation.Although an exact root cause could not be determined a potential root cause could be material selection.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "¿care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation.¿the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.Directions for use: 1.Determine the proper stent length for the patient.This is generally calculated from the baseline 2.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.3.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.4.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.5.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).6.Withdraw the guidewire slowly.The stent will form a pigtail automatically.7.Carefully remove the push catheter." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during placement of the urethral stent by the director and professor on july 21st, 2023, the product was unpacked after the inspection showed that the outer package was intact.The guide wire could not be inserted.The head nurse was asked to use another new box of the urethral stent to complete the procedure.
|
|