Model Number M00570740 |
Device Problems
Break (1069); Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0501 captures the reportable event of inner sheath detached.
|
|
Event Description
|
It was reported to boston scientific corporation on (b)(6) 2023, that a wallflex biliary rx partially covered stent was to be implanted in the common bile duct to treat a malignant biliary stricture during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the handle broke, the inner sheath snapped inside the catheter, and the stent could not be fully deployed.The stent was removed from the patient partially deployed on the delivery system.The procedure was completed with another wallflex biliary stent.There were no patient complications reported as a result of this event.
|
|