Catalog Number DP-SDP001
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Obstruction/Occlusion (2422)
Event Date 07/16/2023
Event Type  Injury  
Event Description
Per rep - (b)(6) 2023 - the clip that was securing the doppler to the vein slipped down the vein and occluded the blood supply.The patient originally had surgery on (b)(6) 2023.The patient had to return to the o.R.On (b)(6) 2023 due to the occluded blood supply.The cause of the occluded vein was determined during the return to the o.R.
 
Manufacturer Narrative
Fill in the common name/product code/510k information if needed, ¿this report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.The event is currently under investigation.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available or upon completion of investigation.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Per rep - 07aug2023 - the clip that was securing the doppler to the vein slipped down the vein and occluded the blood supply.The patient originally had surgery on (b)(6) 2023.The patient had to return to the operating room on (b)(6) 2023 due to the occluded blood supply.The cause of the occluded vein was determined during the return to the operating room.
 
Manufacturer Narrative
G5 ¿ pma/510(k): k17122.The device was disposed of by the customer per tw.The device was not returned for the complaint; therefore, a physical investigation could not be performed, and the customer's complaint could not be confirmed, other than by the customer's testimony.The complaint/event that was entered and reported within trackwise: "the clip that was securing the doppler to the vein slipped down the vein and occluded the blood supply." the device history record (dhr) was reviewed.There were no signs that the device was nonconforming or that the device was not manufactured to current specifications.This complaint mode will be tracked, trended and monitored in cvi complaint handling and post market surveillance procedures.A risk assessment will be performed within the complaint summary tab of trackwise.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK-SWARTZ DOPPLER PROBE
Type of Device
ITX TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
COOK VANDERGRIFT INC
1186 montgomery lane
vandergrift PA 15690
Manufacturer (Section G)
COOK VANDERGRIFT INC
1186 montgomery lane
vandergrift PA 15690
Manufacturer Contact
timothy vogel
1186 montgomery lane
vandergrift, PA 15690
7248458621
MDR Report Key17486661
MDR Text Key321524276
Report Number2522007-2023-00011
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00827002213630
UDI-Public(01)00827002213630(17)251031(10)N191902
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberDP-SDP001
Device Lot NumberN191902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age52 YR
Patient SexFemale
-
-