Dermatome (id 3539250) was returned for evaluation.Failure analysis - the dermatome ran with ample power (measuring.5 amps).During evaluation the dermatome was run multiple times for a minute without skipping.Motor measured.35 amps without the load of the motor casting assembly.Head assembly calibration was found to be in tolerance, measuring cap-0.0021, and bushing -0.0024.Head component had minor nicks and scratches that would be considered normal for the age of the device.Wear marks from the blade around the oscillating hole is older, the area is smooth and free of burrs.Handle assembly found to be in good working condition.Internal handle assemblies were replaced with last repair and remain in good condition.No assembly defects were found.Root cause analysis- unit was found to be within tolerance/specification.No issues were found, and it functions as intended.A possible root cause of the reported failure could be a result of attaching the width clip too tight, causing the blade to pinch between the blade bed and the clip bottom.It is also possible the width clips are worn and out of flatness specification.Width clips were not received with this or any recent returns, so this issue cannot be fully confirmed.
|