Model Number 865350
Device Problems No Device Output (1435); No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2023
Event Type  malfunction  
Event Description
The customer reported that there was no sound from the speaker at all, even during boot up.A speaker inop message appeared.There was no patient involvement.
 
Manufacturer Narrative
Philips is in process of obtaining additional information.A final report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the speaker produced sound.Based on the information available and the testing conducted, the reported problem was not confirmed.Although the speaker was confirmed to be functioning per specification during testing it was indicated that the device is displaying a speaker malfunction inop message and no sound at the time of the event, the speaker has been replaced per current process.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17490280
MDR Text Key320732425
Report Number1218950-2023-00553
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082236
UDI-Public00884838082236
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-