H4: the suspected lot # 60439522 was manufactured from february 24, 2023, to february 28, 2023.H4: the suspected lot # 60438225 was manufactured from february 20, 2023, to february 23, 2023.H10: the actual device was received for evaluation.Visual inspection was performed and valve set port 20 was observed with valve core damaged and was incorrectly oriented.Functional testing and duplication of the issue could not be performed due to the damage and incorrect orientation.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted for the suspected lot numbers 60439522 and 60438225, and there were no deviations found related to this reported condition during the manufacture of these lots.Should additional relevant information become available, a supplemental report will be submitted.
|