Catalog Number H938724
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2023
Event Type  malfunction  
Event Description
It was reported that valve 20 of an em2400 valve set was cracked and was drawing air.This was identified before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
D4: lot #: suspect lots 60439522 and 60438225 were reported.E1: initial reporter address: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the suspected lot # 60439522 was manufactured from february 24, 2023, to february 28, 2023.H4: the suspected lot # 60438225 was manufactured from february 20, 2023, to february 23, 2023.H10: the actual device was received for evaluation.Visual inspection was performed and valve set port 20 was observed with valve core damaged and was incorrectly oriented.Functional testing and duplication of the issue could not be performed due to the damage and incorrect orientation.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted for the suspected lot numbers 60439522 and 60438225, and there were no deviations found related to this reported condition during the manufacture of these lots.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17492447
MDR Text Key321235838
Report Number1416980-2023-03989
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-