• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO UPPSALA AB HEALON 5 PRO; AID, SURGICAL, VISCOELASTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO UPPSALA AB HEALON 5 PRO; AID, SURGICAL, VISCOELASTIC Back to Search Results
Model Number TV60ML
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/18/2023
Event Type  malfunction  
Manufacturer Narrative
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a - implant date: not applicable.Healon is not an implantable device.Section d6b - explant date: not applicable.Healon is not an implantable device.Section e1 - telephone number:(b)(6).Section h3 - other (81): the device was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible device return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that a piece of white foreign material was observed entering the eye upon use of healon5 pro.The customer said this has happened a number of times in the past but was never reported by them.The surgeon managed to aspirate the piece of material which delayed the procedure by 2 minutes.The patient was reported as fully recovered.No further information was provided.A separate report is being submitted for the unknown number of unreported events.
 
Manufacturer Narrative
Additional information: section d9 - device available for evaluation? yes section d9 - date returned to manufacturer: 20-sep-2023 section h3 - device evaluated by manufacturer? yes device evaluation: analysis of the photograph provided by the customer revealed a white speck in the eye.The foreign material found during the operation was received for investigation.The complaint product was not included in the shipment.The material had been placed in a sterile container.Visual inspection and fourier transform infrared spectroscopy (ftir) micro-spectroscopic analysis of this material was carried out.The sample spectrum was compared to a reference infrared (ir) database for identification of the material.The material was spectrally similar to a material identified as ¿rubber membrane material".Based upon the complaint file picture, visual & stereomicroscopic observations, the ftir investigation and the customer narrative, the device problem code dc-foreign material loose was confirmed.The foreign material was spectrally and physically similar to the rubber perforation membrane or the rubber plunger.Coring of the perforation membrane during activation of the cylinder was the most likely source of the particle.This risk is indicated in the product's directions for use (dfu) as follows "express a small amount of the healon5® pro ovd from the syringe prior to use and carefully examine it during use to avoid injecting minute rubber particles which may be released when the syringe diaphragm is punctured".Conclusion: as a result of the investigation, there was no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALON 5 PRO
Type of Device
AID, SURGICAL, VISCOELASTIC
Manufacturer (Section D)
AMO UPPSALA AB
rapsgatan 7
uppsala, uppsala län 754 5 0
SW  754 50
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key17492546
MDR Text Key321107642
Report Number3012236936-2023-01943
Device Sequence Number1
Product Code LZP
UDI-Public(01)(17)250731(10)UK31416
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P810031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTV60ML
Device Catalogue Number30410400
Device Lot NumberUK31416
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-