Model Number HL-390 |
Device Problems
Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other text: b3: date of event and d4: udi number is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the fluid warmer kept "popping" circuit breakers.Discovered during preventative maintenance.No patient involvement was reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Per visual inspection cracked enclosure, corroded drain fitting, and worn line cord.Outdated printed circuit board (pcb) and power switch.Per functional testing the device failed the electrical test because the voltage put out from the device was too high confirming the complaint.The root cause was a faulty pcb.It was unknown what caused it to become faulty.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
|
|