The field application specialist (fas) confirmed that on the day of the event, there was an issue with the result sequence and the result transmission.No data or actual results were provided for the investigation.Based on other information provided, the investigation noted there were some instrument issues when placing the test strip onto the test strip tray.The field service engineer (fse) replaced the printed circuit boards (pcb) for the test strip detector and the strip check.After the service visit, the customer had no further issues.As no specific data was provided for the investigation, the cause of the event could not be determined.
|