Model Number 0672
Device Problems Pocket Stimulation (1463); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 07/31/2023
Event Type  Injury  
Event Description
It was reported that the patient implanted with this right ventricular (rv) defibrillation lead presented to the emergency room four days post implant with complaint of shocks.Review of stored device data showed no delivered shocks, however, the rv lead exhibited high threshold measurements and sensing values were noted to have decreased.An x-ray was performed and confirmed the rv lead had dislodged.Surgical intervention was undertaken two days later.The rv lead was successfully repositioned.No additional adverse patient effects were reported.This device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17494940
MDR Text Key320750670
Report Number2124215-2023-42463
Device Sequence Number1
Product Code LWS
Combination Product (y/n)Y
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0672
Device Catalogue Number0672
Device Lot Number213694
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-