COCHLEAR LIMITED NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI622
Device Problems Circuit Failure (1089); High impedance (1291)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 06/22/2023
Event Type  malfunction  
Event Description
Per the clinic, the patient experienced no sound, increased impedances and open circuits.Troubleshooting attempts were made; however, the issue could not be resolved.The device was explanted on (b)(6) 2023, and the patient was re-implanted with another cochlear device during the same surgery.
 
Manufacturer Narrative
This report is submitted on august 11, 2023.
 
Manufacturer Narrative
Device analysis indicated device failure.Device analysis report attached.This report is submitted on august 24, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI622 COCHLEAR IMPLANT WITH SLIM STRAIGHT ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
hemavathi muniesparan
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key17505538
MDR Text Key320853343
Report Number6000034-2023-02552
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036573
UDI-Public(01)09321502036573(11)230118(17)250117
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCI622
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received08/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexMale
-
-