• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number ASKU
Device Problems Disconnection (1171); Filling Problem (1233); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified elastomeric device leaked and underinfused medication to the patient; further described as "infusion leaking and line disconnected itself" and "device looked full after 18hours of being connected to the 24hr infusion".The nurse checked the device and found the device was still over a half full even with no clamps attached to the bottle.The device was filled with piperacillin/tazobactam.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which revealed that the device contained fluid inside the bladder.Due to the nature of the sample, no additional tests were performed.Based on the photo, the reported condition of under infusion was verified, however, the condition of leak could not be verified by the photograph.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17509010
MDR Text Key321167991
Report Number1416980-2023-04036
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/04/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PIPERACILLIN/TAZOBACTAM
Patient SexFemale
-
-