The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - investigation of the returned unit received has little to no movement in the lock.When the unit had the pins inserted and was placed on the block, there was no movement or slipping.The pressure was pushed up to 80lbs and still could not get the unit to slip.No failure of the device could be duplicated.Root cause - evaluation found no device deficiencies that would have contributed to the reported complaint.When the clamp was properly positioned and put under pressure, it would not move.Probable root cause is improper or suboptimal placement of the skull clamp on the patient.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|
A facility reported that during a posterior cervical procedure, patient was supine on the stretcher and three pin headrest was applied.Patient was flipped prone onto the operating room (or) table.Once prone, the surgeon used the mayfield modified skull clamp (a1059) to achieve flexion and the patient's head slipped in the skull clamp.As a result, laceration occurred and was promptly treated, and the surgery continued with no further complications related to the skull clamp.A delay of 10 minutes occurred due to product problem.
|