• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053 Back to Search Results
Catalog Number D134722IL
Device Problem Complete Blockage (1094)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/06/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.A visual inspection, and irrigation test of the returned device were performed in accordance with bwi procedures.Visual analysis revealed reddish material inside the pebax.The device failed the cool flow pump test as high-pressure values were observed, the device was dissected, and the irrigation tube was found bent in the shaft area, this failure was confirmed to be the root cause of the high-pressure values observed during the test.The pebax was microscopically inspected and a hole on its surface was found, it was determined that this issue is unrelated to the event described by the customer, the root cause of this condition may be related to manipulation of the device; however, this could not be conclusively determined.A manufacturing record evaluation was performed for the finished device 30970245l, and no internal action was found during the review.The issue reported by the customer was confirmed.Product failure is multifactorial.The instructions for use contain the following recommendations: flush the catheter with heparinized saline prior to insertion into the body.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.E1.Initial reporter facility name: (b)(6) hospital.E1.Initial reporter city: (b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.(b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter for which biosense webster¿s product analysis lab (pal) identified a hole on the surface of the device.It was initially reported by the customer that that during the operation, device (including port, luer hub) was not irrigating.When connecting the ablation catheter and performing routine exhaust bubbles, it was found that the blowing nozzle did not spray water.The correct catheter settings were selected on the generator and the pump was switching from low to high flow during ablation.A second device was used to complete the operation.There was no adverse event reported on patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
SIMILAR DEVICE D133601, PMA # P030031/S053
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17511805
MDR Text Key321158697
Report Number2029046-2023-01761
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/10/2024
Device Catalogue NumberD134722IL
Device Lot Number30970245L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Date Manufacturer Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-